The JAL-Z colorimeter is a compact analytical instrumentation for online analysis and process monitoring. The colorimeter is based on a color-developing chemical reaction carried out inside the analyzer's glass cell. The sample, coming from an external reservoir or process stream is autonomously grabbed by the instrument and mixed with the appropriate reagents to make the reaction happen. The analyzer uses an LED light source and a photodiode to photometrically measure the developed color intensity (absorbance). The absorbance is proportional to the concentration of the target chemical species (analite).
The determination is based on the reaction of 1,5-diphenylcarbazide with Chromium(VI) in an acid medium. The absorbance intensity is proportional to the chromium concentration in the sample and is measured at 525 nm.
Measured parameter: Cr(VI)
Range :
(Ø 26 mm) 0 - 300 ppb
(Ø 16 mm) 0 - 1000 ppb
(with dilution) 40 ppm
Reproducibility*
(Ø 26 mm) ± 1 ppb / ± 5% (0-50 ppb),
± 2 ppb / ± 5% (50-300 ppb)
(Ø 16 mm) ± 5 ppb / ± 5% (≥ 300 ppb)
* if two values take whichever is greater
Limit of detection : 0.05 ppb
Wavelength : 525 nm
Demineralized water : Not required
Reference : ISO 11083:1994
Number of reagents: 2
Min. analysis time : 8 minutes
Default analysis time : 15 minutes
Reagent consumption**
(Ø 26 mm) 1.5 L
(Ø 16 mm) 1 L
** per month, at the default analysis frequency.
The determination is based on the reaction of 1,5-diphenylcarbazide with Chromium(VI) in an acid medium. The absorbance intensity is proportional to the chromium concentration in the sample and is measured at 525 nm.
Measured parameter: Cr(VI)
Range :
(Ø 26 mm) 0 - 600 ppb
(Ø 16 mm) 0 - 2 ppm
(with dilution) 80 ppm
Wavelength : 572 nm
Demineralized water : Not required
Number of reagents: 2
Min. analysis time : 8 minutes
Default analysis time : 15 minutes
Reagent consumption**
(Ø 26 mm) 2 L
(Ø 16 mm) 1 L
** per month, at the default analysis frequency.
Chromium in the +3 oxidation state is oxidized to chromium(VI). All the chromium, now as Cr(VI) then reacts with 1,5-diphenylcarbazide, producing a pink color. The intensity of the color is proportional to the total chromium concentration in the sample and is measured at 525 nm.
Measured parameter: Cr(III)+Cr(VI)
Range :
(Ø 26 mm) 2 ppm
(Ø 16 mm) 4 ppm
(with dilution) 150 ppm
Wavelength : 572 nm
Demineralized water : Not required
Number of reagents: 3
Min. analysis time : 11 minutes
Default analysis time : 15 minutes
JAL-Z_Chlorine-Colorimeter-Datasheet (pdf)