The JAL-Z colorimeter is a compact analytical instrumentation for online analysis and process monitoring. The colorimeter is based on a color-developing chemical reaction carried out inside the analyzer's glass cell. The sample, coming from an external reservoir or process stream is autonomously grabbed by the instrument and mixed with the appropriate reagents to make the reaction happen. The analyzer uses an LED light source and a photodiode to photometrically measure the developed color intensity (absorbance). The absorbance is proportional to the concentration of the target chemical species (analite).
The determination ranges of the cyanide analyzer vary from trace μg/L to 10 mg/L CN- using the (optional) internal dilution module.
The cyanide ions present in the sample reacts with the chloramine-T and pyridine/barbituric acid reagents. The absorbance intensity is proportional to the cyanide concentration in the sample and is measured at 572 nm.
Measured parameter: CN-
Range :
(Ø 26 mm) 2 - 100 ppb
(Ø 16 mm) 10 - 200 ppb
(with dilution) 15 ppm
Reproducibility*
(Ø 26 mm) ± 4 ppb / ± 5%
(Ø 16 mm) ± 10 ppb / ± 5%
* if two values take whichever is greater
Limit of detection : 2 ppb
Wavelength : 572 nm
Demineralized water : Not required
Number of reagents: 3
Min. analysis time : 13 minutes
Default analysis time : 20 minutes
Reagent consumption**
(Ø 26 mm) 2 L
(Ø 16 mm) 1 L
** per month, at the default analysis frequency.