The JAL-Z colorimeter is a compact analytical instrumentation for online analysis and process monitoring. The colorimeter is based on a color-developing chemical reaction carried out inside the analyzer's glass cell. The sample, coming from an external reservoir or process stream is autonomously grabbed by the instrument and mixed with the appropriate reagents to make the reaction happen. The analyzer uses an LED light source and a photodiode to photometrically measure the developed color intensity (absorbance). The absorbance is proportional to the concentration of the target chemical species (analite).
The determination ranges of the nickel analyzer vary from trace μg/L to 200 mg/L using the (optional) internal dilution module.
In the presence of an oxidizing agent, nickel ions react with dimethylglyoxime in an alkaline solution to form an orange-brown colored complex. The absorbance intensity is proportional to the nickel concentration in the sample and is determined at 470 nm.
Measured parameter: Ni2+
Range :
(Ø 26 mm) 0.01 - 3 ppm
(Ø 16 mm) 0.02 - 6 ppm
(with dilution) 200 ppm
Reproducibility*
(Ø 26 mm) ± 10 ppb / ± 5%
(Ø 16 mm) ± 30 ppb / ± 5%
* if two values take whichever is greater
Limit of detection : 0.01 ppm
Wavelength : 470 nm
Demineralized water : Not required
Number of reagents: 3
Min. analysis time : 8 minutes
Default analysis time : 15 minutes
Reagent consumption**
(Ø 26 mm) 2.5 L
(Ø 16 mm) 2 L
** per month, at the default analysis frequency.
JAL-Z_Chlorine-Colorimeter-Datasheet (pdf)