The JAL-Z colorimeter is a compact analytical instrumentation for online analysis and process monitoring. The colorimeter is based on a color-developing chemical reaction carried out inside the analyzer's glass cell. The sample, coming from an external reservoir or process stream is autonomously grabbed by the instrument and mixed with the appropriate reagents to make the reaction happen. The analyzer uses an LED light source and a photodiode to photometrically measure the developed color intensity (absorbance). The absorbance is proportional to the concentration of the target chemical species (analite).
The instrument is ideal for potable water or suface water analysis and with the addition of a dilution module is also capable of the determination of nitrates in complex matrices such as urban and industrial wastewaters.
The determination ranges of the nitrate analyzer vary from trace μg/L to 25.0 mg/L N-NO3 (80 mg/L NO3) using the (optional) internal dilution module. The result can be expressed as NO3 or N-NO3.
Both hydrazine and copper catalyst reagents are used to reduce Nitrate to Nitrite in a basic medium inside a heated reaction cell. A sulfanilamide and N-(1-naphthyl)-ethylene diammonium dichloride reagent is added to react with the Nitrite to form a purple colored compound. The absorbance intensity is proportional to the Nitrite concentration in the sample and is measured at 525 nm
Measured parameter: NO3-
Range :
(Ø 26 mm) 2 - 400 ppb N-NO3 (1.28 ppm NO3)
(Ø 16 mm) 4 - 650 ppb N-NO3 (2.08 ppm NO3)
(with dilution) 25 ppm N-NO3 (80 ppm NO3)
Reproducibility*
(Ø 26 mm) ± 2 ppb / ± 5%
(Ø 16 mm) ± 5 ppb / ± 5%
* if two values take whichever is greater
Limit of detection : 2 ppb
Wavelength : 525 nm
Demineralized water : Not required
Termoregulation : 37 °C
Interferences : Amines, urea, iodides.
Reference : DIN EN 26 777
Number of reagents: 3
Min. analysis time : 11 minutes
Default analysis time : 20 minutes
Reagent consumption**
(Ø 16 mm) 4 L
** per month, at the default analysis frequency.
JAL-Z_Chlorine-Colorimeter-Datasheet (pdf)