The JAL-Z colorimeter is a compact analytical instrumentation for online analysis and process monitoring. The colorimeter is based on a color-developing chemical reaction carried out inside the analyzer's glass cell. The sample, coming from an external reservoir or process stream is autonomously grabbed by the instrument and mixed with the appropriate reagents to make the reaction happen. The analyzer uses an LED light source and a photodiode to photometrically measure the developed color intensity (absorbance). The absorbance is proportional to the concentration of the target chemical species (analite).
The determination ranges of the nitrite analyzer vary from trace μg/L to 125 mg/L NO2 using the (optional) internal dilution module. The result can be expressed as NO2 or N-NO2.
In acidic media, nitrites react with sulfanilamide and N-(1-naphthyl)-ethylene diammonium dichloride, forming a purple colored dye. The absorbance intensity is proportional to the nitrites concentration in the sample and is determined at 525 nm.
Measured parameter: NO2-
Range :
(Ø 26 mm) 5 - 500 ppb N-NO2 (1.6 ppm NO2)
(Ø 16 mm) 0.02 - 1 ppm N-NO2 / (3.2 ppm NO2)
(with dilution) 40 ppm N-NO2 / 125 ppm NO2
Reproducibility*
(Ø 26 mm) ± 5 ppb / ± 5%, (0-150 ppb)
(Ø 16 mm) ± 10 ppb (150-600 ppb), ± 20 ppb / ± 5% (≥ 600 ppb)
* if two values take whichever is greater
Limit of detection : 5 ppb
Wavelength : 525 nm
Demineralized water : Not required
Interferences : Amines, urea, iodides.
Reference : DIN EN 26 777
Number of reagents: 2
Min. analysis time : 7 minutes
Default analysis time : 15 minutes
Reagent consumption**
(Ø 26 mm) 1.5 L
(Ø 16 mm) 1 L
** per month, at the default analysis frequency.