The JAL-Z colorimeter is a compact analytical instrumentation for online analysis and process monitoring. The colorimeter is based on a color-developing chemical reaction carried out inside the analyzer's glass cell. The sample, coming from an external reservoir or process stream is autonomously grabbed by the instrument and mixed with the appropriate reagents to make the reaction happen. The analyzer uses an LED light source and a photodiode to photometrically measure the developed color intensity (absorbance). The absorbance is proportional to the concentration of the target chemical species (analite).
The determination ranges of the silica analyzer vary from trace μg/L to 150 mg/L using the (optional) internal dilution module.
Soluble silica reacts with the molybdate ion in an acid medium to form a green-yellow colored silico-molybdic acid complex that in its turn is converted to a blue complex with the addition of the reducing reagent. Oxalic acid is added to minimize the phosphate interference. The absorbance intensity is proportional to the silica concentration in the sample and is determined at 850 nm.
Measured parameter: SiO2
Range :
(Ø 26 mm) 0.5 - 1000 ppb
(Ø 16 mm) 1 - 5000 ppb
(with dilution) 0.05 - 150 ppm
Reproducibility*
(Ø 26 mm) ± 1 ppb / ± 5%
(Ø 16 mm) ± 1.5 ppb / ± 5%
* if two values take whichever is greater
Limit of detection : 0.5 ppb
Wavelength : 850 nm
Demineralized water : Not required
Number of reagents: 4
Min. analysis time : 10 minutes
Default analysis time : 15 minutes
Reagent consumption**
(Ø 26 mm) 2.9 L
(Ø 16 mm) 1.5 L
** per month, at the default analysis frequency.
JAL-Z_Chlorine-Colorimeter-Datasheet (pdf)